Evaluasi Kontrol Kualitas Pemeriksaan Glukosa Darah di Laboratorium X Palangka Raya
Evaluation of Quality Control of Blood Glucose Examination in Laboratory X Palangka Raya
DOI:
https://doi.org/10.33084/bjmlt.v5i2.5184Keywords:
Quality Control, total error, six-sigma, Westgard Multirule SystemAbstract
Quality Control is a managerial procedure to adjust the laboratory examination (analytic) stages and detect errors early. The purpose of this study was to describe the quality control in laboratory X in the city of Palangka Raya. The method used is the descriptive method. The results showed that in April 2022, the Mean (88,1 mg/dL), SD (1,09 mg/dL), CV (1,24%), d (0,14%), TE (2,62%), TEa (10%), Sigma (7,94), and in May 2022 the Mean (87,9 mg/dL), SD (1,00 mg/dL), CV (1,13%), d (0,04%), TE (2,31%), TEa (10%), Sigma (8,79). According to the Ministry of Health, the maximum CV value is < 5%, so the inspection results of control materials in April 2022 and May 2022 are good or thorough. The maximum d value according to ISO 15179 is < 5%, so the inspection results of control materials in April 2022 and May 2022 are good or accurate. The evaluation of the control material with the Levey Jennings chart showed that all values in April 2022 and May 2022 were within the control limits, and none were outside the limits of the Westgard Mulrirule System rules. The result shows that the internal quality control activities of clinical chemistry for glucose parameters in Laboratory X are well executed and reliable.
Downloads
References
Fuadi, R., 2019. Using Six Sigma To Evaluate Analytical Performance Of Hematology Analyzer. Indonesian Journal of Clinical Pathology and Medical Laboratory, 25 (2) : 165 – 169
Geto, Z., Getahun, T., Lejisa, T., Tolcha, Y., Bikila, D., Bashea, C., Meles, M., Habtu, W., Ashebir, G., Negasa, B., Sileshi, M., Daniel, Y., Gashu, A., & Challa, F. 2022. Evaluation of Sigma Metrics and Westgard Rule Selection and Implementation of Internal Quality Control in Clinical Chemistry Reference Laboratory, Ethiopian Public Health Institute. Indian Journal of Clinical Biochemistry, 37(3), 285–293.
Harr, K. E., Flatland, B., Nabity, M., & Freeman, K. P. 2013. ASVCP guidelines: Allowable total error guidelines for biochemistry. Veterinary Clinical Pathology, 42(4), 424–436.
Kumar, B. V., & Mohan, T. 2018. Sigma metrics as a tool for evaluating the performance of internal quality control in a clinical chemistry laboratory. Journal of Laboratory Physicians, 10(02), 194–199.
Kusmiati, M., Nurpalah, R. dan Restaviani, R. 2022. Presisi dan Akurasi Hasil Quality Control pada Parameter Pemeriksaan Glukosa Darah di Laboratorium Klinik Rumah Sakit X Kota Tasikmalaya. Journal of Indonesian Medical Laboratory and Science. 3(1): 27-37.
Mengko, Richard. 2013. Instrumentasi Laboratorium Klinik. ITB: Bandung
Menteri Kesehatan RI. 2013. Peraturan Menteri Kesehatan RI No.43 Tentang Cara Penyelenggaraan Laboratorium Klinik yang Baik. Menteri Kesehatan RI: Jakarta
Oosterhuis, W. P., Bayat, H., Armbruster, D., Coskun, A., Freeman, K. P., Kallner, A., Koch, D., Mackenzie, F., Migliarino, G., Orth, M., Sandberg, S., Sylte, M. S., Westgard, S., & Theodorsson, E. 2018. The use of error and uncertainty methods in the medical laboratory. Clinical Chemistry and Laboratory Medicine, 56(2), 209–219.
Putra, I.G.P.A.F.S., Prihatiningsih, D. dan Subhaktiyasa, P.G. 2020. Analisis Pemantapan Mutu Internal Pemeriksaan Trombosit di Laboratorium Klinik UPTD. Puskesmas Abiansemal I. Jurnal Analis Laboratorium Medik. 5(2): 28-34
Shah Goel, D. S., Saini, D. R., Singh, D. S. B., Aggarwal, D. O., & Goel, D. A. K. 2014. Six Sigma Metrics and Quality Control in Clinical Laboratory. International Journal of Medical Research and Review, 2(2), 140–149.
Westgard, J. O., & Westgard, S. A. 2017. Measuring Analytical Quality: Total Analytical Error Versus Measurement Uncertainty. In Clinics in Laboratory Medicine (Vol. 37, Issue 1, pp. 1–13). W.B. Saunders.
Westgard, S., Bayat, H., & Westgard, J. O. 2018. Analytical sigma metrics: A review of six sigma implementation tools for medical laboratories. In Biochemia Medica (Vol. 28, Issue 2 Special Issue). Biochemia Medica, Editorial Office.
Zulkifli, I.H. 2020. Ketelitian dan Evaluasi Grafik Kontrol Levey Jennings Pemeriksaan Kadar Asam Urat menggunakan Pooled Sera. Skripsi. Universitas 'Aisyiyah Yogyakarta, Yogyakarta.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 Ferina Yudita, Dwi Purbayanti, Fitria Hariati Ramdhani, Eka Jaya
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
All rights reserved. This publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording.